ACA2014: Difference between revisions
Line 3: | Line 3: | ||
== Thaumatin == | == Thaumatin == | ||
1800 frames collected on a Pilatus 6M detector at APS 24-ID-C. What is special about these data: | 1800 frames collected on a Pilatus 6M detector at APS 24-ID-C; data at [ftp://turn5.biologie.uni-konstanz.de/pub/datasets/ACA2014_thaumatin/]. What is special about these data: | ||
* tiny spots (1-2 pixel) | * tiny spots (1-2 pixel) | ||
* shutterless and fast | * shutterless and fast |
Revision as of 12:54, 28 July 2016
The following datasets were used at the "Data processing with the Pros" session of the ACA2014. Keitaro Yamashita's talk about XDS processing may be downloaded.
Thaumatin
1800 frames collected on a Pilatus 6M detector at APS 24-ID-C; data at [1]. What is special about these data:
- tiny spots (1-2 pixel)
- shutterless and fast
- the header is wrong: the direct beam position is at 1288.9 1262.6 rather than at 1265.47 1273.87 . If not corrected, this may lead to mis-indexing
The XDS data processing is documented at Thaumatin_ACA2014.
Trypsin
7200 frames collected (and 3600 frames provided) on a Pilatus 2M detector at BL-1A, PF. What is special about these data:
- shutterless and fast
- low energy (2.7 A wavelength)
The XDS data processing is documented at Trypsin_ACA2014.
Trypsin with a tight-binding protein inhibitor
226 frames collected on a MarCCD 300 detector at BL XXX. What is special about these data:
- hard-to-identify space group
The XDS data processing is documented at Trypsin_inhibitor_ACA2014. Data are at [2].
"Tough Spots"
Frames collected on an ADSC Q315 CCD detector at BL XXX. Only 10 frames are available [3]. What is special about these data:
- streaks rather than "spots"
The XDS data processing is documented at Tough_Spots_ACA2014.
CagL
450 frames collected on an ADSC Q315 CCD detector at BL XXX. What is special about these data:
The XDS data processing is documented at CagL_ACA2014.