  THE "DATA_RANGE=" IN FILE "XDS.INP" AND START ALL OVER AGAIN.
  THE "DATA_RANGE=" IN FILE "XDS.INP" AND START ALL OVER AGAIN.


This is printed out for you to actually read, and take action accordingly. In most cases you just change the JOB= line in XDS.INP to read
This is printed out for you to actually read, and take action accordingly: it may result from one or more additional crystals contributing to the diffraction patterns. Or it may just be the case that there are ice rings with very many isolated ice reflections (in that case you may want to use EXCLUDE_RESOLUTION_RANGE for the IDXREF task, or just to ignore the message and go on with JOB=DEFPIX INTEGRATE CORRECT). Or you have MINIMUM_NUMBER_OF_PIXELS_IN_A_SPOT of 3, but your reflections comprise many more pixels and COLSPOT may "see" several reflections when there's only one (in that case use MINIMUM_NUMBER_OF_PIXELS_IN_A_SPOT=6 or even 12). In most cases you just change the JOB= line in XDS.INP to read
  JOB= DEFPIX INTEGRATE CORRECT
  JOB= DEFPIX INTEGRATE CORRECT
and then continue to run XDS. In other cases you may want to change SPOT.XDS, or other keywords in [[XDS.INP]] (see also below). But in any case this is an important alert that should make you check the correctness of the parameters that describe the data collection (X-RAY_WAVELENGTH, DETECTOR_DISTANCE, ORGX, ORGY, OSCILLATION_RANGE, NAME_TEMPLATE_OF_DATA_FRAMES).
and then continue to run XDS. In other cases you may want to change SPOT.XDS, or other keywords in [[XDS.INP]] (see also below). But in any case this is an important alert that should make you check the correctness of the parameters that describe the data collection (X-RAY_WAVELENGTH, DETECTOR_DISTANCE, ORGX, ORGY, OSCILLATION_RANGE, NAME_TEMPLATE_OF_DATA_FRAMES).
2,684

edits