Beamline notes: Difference between revisions
Jump to navigation
Jump to search
(BM30A) |
(update Diamond ID24) |
||
Line 6: | Line 6: | ||
* BM14 (ESRF; Mar CCD, as of April 2014) has X and Y reversed in the header. [[Generate_XDS.INP]] does not currently implement this reversal; it has to be done manually. | * BM14 (ESRF; Mar CCD, as of April 2014) has X and Y reversed in the header. [[Generate_XDS.INP]] does not currently implement this reversal; it has to be done manually. | ||
* BM30A (ESRF, ADSC) has ROTATION_AXIS=-1 0 0 (implemented in [[generate_XDS.INP]] on 2016-10-02) | * BM30A (ESRF, ADSC) has ROTATION_AXIS=-1 0 0 (implemented in [[generate_XDS.INP]] on 2016-10-02) | ||
* Diamond ID24 | * Diamond ID24 (PILATUS3 6M, S/N 60-0119) moved from horizontal to vertical spindle for cryo data collection. If the CBF header has # Chi 0.0000 deg. and # Chi_increment 0.0000 deg. and # Oscillation_axis X.CW +SLOW then one needs ROTATION_AXIS=0 -1 0 . | ||
* PETRA P14 also needs ROTATION_AXIS=0 -1 0 | * PETRA P14 also needs ROTATION_AXIS=0 -1 0 | ||
* ALS 8.2.2 recent (Nov 2015) evidence indicates that the same beam center convention is used as for ESRF, PF, APS Argonne BLs. | * ALS 8.2.2 recent (Nov 2015) evidence indicates that the same beam center convention is used as for ESRF, PF, APS Argonne BLs. | ||
* Shanghai synchrotron Radiation Facility seems to need ROTATION_AXIS=-1 0 0 | * Shanghai synchrotron Radiation Facility seems to need ROTATION_AXIS=-1 0 0 |
Revision as of 13:57, 6 December 2016
This article is meant to collect information about specific beamlines.
GlobalPhasing's notes are at http://www.globalphasing.com/autoproc/wiki/index.cgi?BeamlineSettings .
- Taiwanese synchrotron (ADSC Q315R, SN 930), 2015 : ROTATION_AXIS=1 0 0 whereas in 2010 it was -1 0 0
- BM14 (ESRF; Mar CCD, as of April 2014) has X and Y reversed in the header. Generate_XDS.INP does not currently implement this reversal; it has to be done manually.
- BM30A (ESRF, ADSC) has ROTATION_AXIS=-1 0 0 (implemented in generate_XDS.INP on 2016-10-02)
- Diamond ID24 (PILATUS3 6M, S/N 60-0119) moved from horizontal to vertical spindle for cryo data collection. If the CBF header has # Chi 0.0000 deg. and # Chi_increment 0.0000 deg. and # Oscillation_axis X.CW +SLOW then one needs ROTATION_AXIS=0 -1 0 .
- PETRA P14 also needs ROTATION_AXIS=0 -1 0
- ALS 8.2.2 recent (Nov 2015) evidence indicates that the same beam center convention is used as for ESRF, PF, APS Argonne BLs.
- Shanghai synchrotron Radiation Facility seems to need ROTATION_AXIS=-1 0 0