Pathologies: Difference between revisions

Line 10: Line 10:
== Hardware problems ==
== Hardware problems ==


=== Scale factor plot in case of spindle problems ===
=== Scale factor plot in case of problems (beam or spindle) ===


[[Image:Scales.png]]
[[Image:Scales.png]]


The scale factor is printed, in INTEGRATE.LP, for every frame (column 3). This plot shows spikes indicating that the spindle went too fast every 13 frames or so (but the spindle went a bit slower in the adjacent frames). Needless to say, this problem is quite detrimental to data quality - the reflections which contribute to those frames that went too fast are underexposed.
The scale factor is printed, in INTEGRATE.LP, for every frame (column 3). This plot shows spikes indicating that the beam was weak, or the spindle went too fast every 13 frames or so (but in that case the spindle went a bit slower in the adjacent frames). No matter what the problem is due to, it is quite detrimental to data quality - the reflections which contribute to those frames that went too fast are underexposed.


=== Mosaicity plots in case of spindle problems ===
=== Mosaicity plots in case of problems ===


[[Image:Sigmar.png]]
[[Image:Sigmar.png]]
2,684