MINIMUM ZETA: Difference between revisions

m
mNo edit summary
Line 81: Line 81:
[[Image:ms688-frame.png]]
[[Image:ms688-frame.png]]


From looking at rf.pck of many datasets, it is my experience that at the SLS (X06SA, Pilatus detector), the R-factors along the spindle are better than perpendicular to it, which is quite surprising (and should be investigated). Therefore it is clear that in particular for these data it is a good thing to decrease MINIMUM_ZETA because accurately measured reflections are added to the data set.
From looking at rf.pck of many datasets, it is my experience that at the SLS (beamline X06SA), the R-factors along the spindle are better than perpendicular to it, which is quite surprising (and should be investigated). Therefore it is clear that in particular for these data it is a good thing to decrease MINIMUM_ZETA because accurately measured reflections are added to the data set.
2,652

edits