It can happen that you have two different mono-crystals in your loop, and that both are in the X-ray beam. If their relative orientation is sufficiently distinct, it is easy with XDS to index and integrate both crystal diffraction from the same data-set. You end-up with two distinct reflection files and can try to scale them using XSCALE to complete or increase the redundancy of your measurement.  
It can happen that you have two different mono-crystals in your loop, and that both are in the X-ray beam. If their relative orientation is sufficiently distinct, it is easy with XDS to index and integrate both crystal diffraction from the same data-set. You end-up with two distinct reflection files and can try to scale them using XSCALE to complete or increase the redundancy of your measurement.  


After indexation and integration of a first lattice, you can extract the un-indexed reflections to create a new SPOT.XDS file (don't forget to copy the result of the first processing!) and re-run XDS from the IDXREF stage :
After indexing and integration of a first lattice, you can extract the un-indexed reflections to create a new SPOT.XDS file (don't forget to copy the result of the first processing!) and re-run XDS from the IDXREF stage :


  mkdir xtal1
  mkdir xtal1
Line 65: Line 65:
  echo " JOB= IDXREF INTEGRATE CORRECT" >> XDS.INP
  echo " JOB= IDXREF INTEGRATE CORRECT" >> XDS.INP
  xds_par
  xds_par
== pick the h+k+l=2n reflections from a primitive dataset ==
grep \! XDS_ASCII.HKL | grep -v "END_OF_DATA" > x 
grep -v \! XDS_ASCII.HKL | awk '{if ( ($1+$2+$3)%2==0 ) print $0}'  >>x
echo \!END_OF_DATA >> x
and now use e.g.
phenix.xtriage x
to analyze x in terms of body-centered statistics.