Finding out ORGX ORGY: Difference between revisions

Jump to navigation Jump to search
Line 22: Line 22:


Second method: if you feel you would rather like to find ORGX, ORGY by smart trial-and-error, then you just need to realize that in the output of IDXREF.LP you can check the lines
Second method: if you feel you would rather like to find ORGX, ORGY by smart trial-and-error, then you just need to realize that in the output of IDXREF.LP you can check the lines
  DETECTOR COORDINATES (PIXELS) OF DIRECT BEAM    361.92    533.30 ! this should agree with XDS-Viewer findings
  DETECTOR COORDINATES (PIXELS) OF DIRECT BEAM    361.92    533.30 ! this should agree with XDS-Viewer findings
  DETECTOR ORIGIN (PIXELS) AT                      512.37    533.63 ! this should agree with specified ORGX, ORGY
  DETECTOR ORIGIN (PIXELS) AT                      512.37    533.63 ! this should agree with specified ORGX, ORGY
i.e. you have to change ORGX, ORGY in XDS.INP to make the upper line show the values you expect from clicking the direct beam position in XDS-Viewer.
2,684

edits

Cookies help us deliver our services. By using our services, you agree to our use of cookies.

Navigation menu