2,684
edits
No edit summary |
|||
Line 132: | Line 132: | ||
JOB= DEFPIX INTEGRATE CORRECT | JOB= DEFPIX INTEGRATE CORRECT | ||
and then continue to run XDS. In other cases you may want to change SPOT.XDS, or other keywords in [[XDS.INP]] (see also below). But in any case this is an important alert that should make you check the correctness of the parameters that describe the data collection (X-RAY_WAVELENGTH, DETECTOR_DISTANCE, ORGX, ORGY, OSCILLATION_RANGE, NAME_TEMPLATE_OF_DATA_FRAMES). | and then continue to run XDS. In other cases you may want to change SPOT.XDS, or other keywords in [[XDS.INP]] (see also below). But in any case this is an important alert that should make you check the correctness of the parameters that describe the data collection (X-RAY_WAVELENGTH, DETECTOR_DISTANCE, ORGX, ORGY, OSCILLATION_RANGE, NAME_TEMPLATE_OF_DATA_FRAMES). | ||
See also [[Problems#frame_numbers_beyond_999999]] below. | |||
=== IDXREF produces too short cell parameter(s) === | === IDXREF produces too short cell parameter(s) === |