Finding out ORGX ORGY: Difference between revisions

No edit summary
Line 23: Line 23:
Second method: if you feel you would rather like to find ORGX, ORGY by smart trial-and-error, then you just need to realize that in the output of IDXREF.LP you can check the lines
Second method: if you feel you would rather like to find ORGX, ORGY by smart trial-and-error, then you just need to realize that in the output of IDXREF.LP you can check the lines


  DETECTOR COORDINATES (PIXELS) OF DIRECT BEAM    361.92    533.30 ! this should agree with XDS-Viewer findings
  DETECTOR COORDINATES (PIXELS) OF DIRECT BEAM    361.92    533.30 ! should be close to XDS-Viewer findings
  DETECTOR ORIGIN (PIXELS) AT                      512.37    533.63 ! this should agree with specified ORGX, ORGY
  DETECTOR ORIGIN (PIXELS) AT                      512.37    533.63 ! should be close to specified ORGX, ORGY


i.e. you have to change ORGX, ORGY in XDS.INP to make the upper line show the values you expect from clicking the direct beam position in XDS-Viewer.
i.e. you have to ''change ORGX, ORGY in XDS.INP to make the upper line show the values you expect'' from clicking the direct beam position in XDS-Viewer.
2,684

edits